Here is the best image obtained so far:
Image of laser fiber focus to focus using the foreoptics ellipsoid alone
The detector here has 5.4µm pixels, so the FWHM across the widest part of the image (yellow cut line) is about 190µm...not sure how this would scale to the detector, maybe someone can comment. This was done with an f/35-ish beam size incident on the ellipsoid.
To get this image we intentionally decentered the ellipsoid by 10mm in the dimension perpendicular to the plane formed by the centers of the foreoptics mirrors and the laser fiber. Dana calls this the Y-axis. With this intentional de-center, running the camera through best focus looks like this:
Decentered focus run (click to see animation)
Focus run w/beam nominally centered on ellipsoid (click to see animation)
There are only three frames is the 'centered' .gif because we did not use as fine a resolution for focus checks, but the plate scale is the same for the two animations, and we get to a smaller spot size when we decenter. Is this an indication of a clocking error? Inquiring minds wish to know.
We can also look at the image as we translate the mirror, though it's harder to make a movie out of since the image moves around on the camera:
Image variation with 'Y-axis' translation of ellipsoid (click to see animation)
Translating in the perpendicular direction (in the plane of the back surface of the ellipsoid, if that makes any sense, we don't see much sensitivity to decenter:
Image variation with 'X-axis' translation of ellipsoid (click to see animation)
We'll need to cogitate over what this may mean, if anything. The foreoptics is now at the position which gives the image at the top of this post. Dana has talked about shifting the clocking and repeating the decentering tests, but it might be well to compare these images to Zemax predictions to guide the next moves.