Below are spectral focus plots of SLIT1. The pinhole spot symmetry is good for all these settings, and I'll provide those data later.
Setting 16/7 on diag 23 is the best setting for the left hand side of the detector. I extended the diagonal to lower echelle turns and it gets worse for 17/6, 18/5 ...
Also went to slit3, 2.7km/s and verified that > 103000 resolution is being attained for this setting.
Will show these data when the rest of the slit spectra are taken.
The best setting for the right hand side of the detector is 15/10 on the diagonal 25. Since it appears I may be able to do better at higher diagonals and higher echelle turns I am taking this data now.
Since our best echelle pinhole spectra are from the right hand side, finding the best right hand focus in high-mode may be more relevant.
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 25 |
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 24 |
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 23 |
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 22 |
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 21 |
![]() |
Spectral FWHM of SLIT1 spectra along Diagonal 20 |