Some analysis tasks related to CD15. When done, there should be a short answer and/or link to a blog posting with detailed results.
- Medium mode focus:
- High mode focus:
- Filter lightleaks:
- Filter throughputs:
....
- Blackbody efficiency (relating photon flux to temperature - look at KRS5 and lab air transmission)
...
- TAAS information (get EXES platescale in arcsec/pixel, EXES f/cone, alignment)
From the updated version of TAAS Summary Post, the FCLS unit is mm and 1 mm = 4.16". The blog posting has that ~21.5 EXES pixels is 1 mm of FCLS motion or 4.16". This would give a platescale of 0.1935"/pixel. Nominal design value is 0.206"/pix for 25 micron pixels and F/10 beam in EXES. This is a 6% difference. Close enough for now.
From CD15, Acceptance angle from TAAS post finds that the acceptance angle is 0.060 +/- .004 radians. The design for the incoming beam from SOFIA is supposed to be f/19.6 or 0.051 radians. Not sure what to make of this. Would like to try imaging 2 soldering irons at distance of secondary from entrance window during CD16
From CD15, Acceptance angle from TAAS post finds that the acceptance angle is 0.060 +/- .004 radians. The design for the incoming beam from SOFIA is supposed to be f/19.6 or 0.051 radians. Not sure what to make of this. Would like to try imaging 2 soldering irons at distance of secondary from entrance window during CD16
- Bounce test (sensitivity to impulsive vibrations when flip mirror against stops and not, motors engaged and not)
See the Sep 18, 2013 Blog posting (bounce-tests-cd015.html) for images. In high-res at aligned position with motors enabled, see no evidence for bounce. Motors disabled see spots move <spot diameter in both spatial and spectral dimension. Disabled against the spot is about the same magnitude, but almost purely in spatial direction. In medium mode, aligned/disabled test looks pretty good, other than the noise.
- Pupil image focus test (may not have worked too well. Mostly care about internal focus - how well do we image the cold-stop cross hair)
...
- Ability to image through different filters (can we image room and/or blackbody without going severely non-linear in select filters)
Checked N04582 and N10247-9. N04582 worked better, but needed partial array readout and very fast to get just below saturation level (5200 instead of 5000).
- Results of shimming ellipsoid and what to do for next cooldown
....
- General light leaks (compare counts with lens wheel on mirror with counts when slit on mirror and with counts in unilluminated regions)
...