First Second pass at CD16 tests. This reflects discussion on 10/22
This is posted mostly so people can comment and we can revise. What else?!?
- would like to run as much as possible from GUI.
- Should have bias we expect to fly with (2 V?). If we need to adjust bias when higher photon fluxes, we can do it, but that would mean we need to get bias board installed for commissioning.
- Would like to run with short frame time as much as possible. ctdelay = 13 will give us 493 msec. Do lots of Ns (and use lots of memory).
- verify focus across array with flat focal plane
- medium mode
- hi-med mode (slit 3 should give < 3 km/s at 14 microns)
- Check for leaks in BBP_4700_5615 + MgF2 blocker.
- Calibration testing - ideal is to be able to do calibrations in the lab before/after mounting to SOFIA, but that probably won't happen. We'd need to be able to use a model for dividing out the sky for this to work. More likely is to do a blackbody and sky at start of each leg for the calibrations with eventual goal of only needing to do it once per flight per setting.
- flatfield - how repeatable are blackbody images. Good to do at wavelength with little/no atmosphere. 1230 cm-1 is one possibility (N07904 filter). Or medium mode long slit near 1160 cm-1 (Ocli Q in decker).
- in time series (differences with no change to setup)
- after lunch or overnight (leave setup as is, but let time change)
- after returning to an old setup
- Note: want to check flatfield counts vs predictions.
- darks (will be subtracted from blackbody):
- how dark is mirror in lens wheel compared to dark at slit wheel and to unilluminated parts of cross dispersed and long slit.
- how variable is it (various time scales).
- Linearization data - repeats of multi-Ns with D such that we can see ADU change as a function of linear increase in time. Vary illumination by using slit width? Have orders just slightly overlapping in hi-med mode.
- Check to see if ellipsoid shimming is good.
- Check pupil imaging focus, if needed.
- Get throughput for filters without them. Also need throughput for at least one previous filter to see if bias has impact.
- Noise data: verify noise floor with new fanout using series of identical darks (lens wheel mirror) at several biases.
- Measure input angle using heat source at distance of secondary mirror. TAAS SCHP with no SI plate (or SI plate still connected to EXES) might work. Might use travel of SCHP to determine acceptance angle. Distance between EXES window and secondary is (from KRS5 zemax file) 6.66 meters or 21 feet 10 inches. Secondary diameter is 352 mm or 13.86 inches. I think we'd like to have SCHP travel be larger than 70 mm (20% of secondary diameter) to get an accurate enough idea of our acceptance angle. More travel is better.
- Mount to TAAS
- Compare alignment to SCHP on TAAS to previous iterations. Ted Dunham says he's seen 1.5mm difference in TAAS FCLS locations. Absolutes are, therefore, questionable. Relative motions should, I think, be ok.
- Look for change of SCHP in cold stop as a function of letting EXES sag on its plate.
4 comments:
One thing that's new for this cooldown is the use of an unmodified fanout card....all previous work has been done with modifications to compensate for missing wires in the instrument cabling. Since these are now fixed, we need to verify our noise floor in this configuration. At minimum we need a series of identical darks (lens-wheel blocked) at several biases.
More likely, we should do this as part of a larger set of tests to populate the instrument performance table with an eye toward demonstrating how well EXES is working during the preship review.
Order of priorities and suggested sequence for tests.
note: EXES was left in OPEN/OPEN/OPEN, echelle=-1.3 deg, high-res
Day 1:
1.1. go to V-bias=2V
1.2. go to a narrow, short filter, tune out odd/even effect (save image)
center up on SDF with echelle.
1.3) go to medium res pupil imaging mode, decker dash 4 x slit 5, filter=N07xxxx (pick lowest flux, d=1" filter)
lens= KRS5
move flip motor roughly -21000 steps (keep count), using caution not to run into wall. walk forward
+50 increments until maximize pupil size and roundness is reached. Save image, record flip voltage
& cumulative steps from initial high-res position of flip.
1.4 go to camera mode OPEN/OPEN/N07xxxx and iteratively change decker and filter position to maximize aperture
center SDF position, and input new v0s into GUI. Record "flip angle" and flip voltage and take an image.
1.5. install cross target on EXES window with mirror in back to check the ellipsoid outlooking angle. take images.
use parallel line cross, 10mm apart. go to pupil lens and take image. (#5)
1.6. go to high-res pupil (KRS5), and walk up +steps until oval maximized. Record voltage and angle and take image. record steps (=N). Without moving lens, go -(N+500) steps, then +500 steps to verify step count to medium mode.
(work with Mike to try mode change with flip from GUI, between steps 8.
1.7. go to lens=open, echelle=56 deg for 4th order Q-branch medium. Get C2H2 lines, and find the screwdriver count
on echelle.
1.8. slit1 focus scan from +6T to +15 T on echelle parabaloid. no rinsing this time.
IDL processing to find best focus position (30 min). (#1.1, partial)
1.9. pinhole focus scan from +6 to +15T on echelle paraboid, with rinses. Leave at best slit1 focus. (#1.1, partial)
1.10. go to high-res using GUI. take a snapshot in pupil mode to make sure same flip position is reached.
find echelon screwdriver count.
1.11. focus scan of echelon parabaloid mirror with slit1 of +7 to +16. (#1.2, partial)
1.12. focus scan of echelon parabaloid mirror with pinholes from +7 to +16, with pinhole rinses.
(while processing the slit data) (#1.2, partial)
End of day: High-res slit data needs to be processed for next day's work. Pinhole aspect ratio should be checked, but further
processing can wait.
Days 2-5
2.0 put echelon drivers at best focus position. Take slit1-6 high-res data with C2H2. Take slit 1-6 data in medium with C2H2.
2.1 return to pupil imaging mode (now that best echelle focus is achieved.) Record images with ZnSe and KRS5 lens. (#6)
2.2 go to high-res pupil imaging with GUI, and check pupil is the same as before.
2.3 move EXES without SI plate up to TAAS with SCHP installed and do acceptance cone measurements (#9).
x Travel on SCHP is ~300mm.
2.4 Put SI plate on.
2.5 BBP_4700 filter scan
2.6 Calibration testing (probably need a little time write evaluation code on the fly and repeat things as needed.
Do filter throughputs, linearization and dark tests during "thinking time" about the calibrations. (#3, #4, #7, #8)
(Week 2)
Day 8
8.1 Mount to TAAS and perform SCHP experiments (#10).
8.2 Jeff Huang needs photos of this step for pre-ship review.
8.3 Remove from TAAS and remove plate.
8.4 Followup focus stuff- did we get good enough focus for slit3, or should we go to another diagonal?
8.5 If no changes to final focus, then do Slit1-slit6 at 6 and 17um for high, medium, high and low resolution modes.
Do 14um slit1-6 for low resoluton mode.
Day 9 +
Mode change repeatability, other mechanism repeatability, return to TAAS to try spectrum of FCLS with offsets for pseudo-nodded data set. Longer term flat fields stability. Efficiency of corner cube modes on low-resolution mode.
1.3 and 1.4) N04679 may have lower flux. It has better atmospheric transmission than the other 4-5 micron filters we had before. I was thinking this would be our best bet for imaging during line ops. N07904 likely will be the next best, again basing this only on atmospheric transmission.
1.3) Could go to lens wheel = ZnSe pupil lens for quick comparison of cold stop focus.
1.5) have we been doing cross target imaging in pupil mode? I thought it was in imaging. Lens wheel to open. S/D/F to open.
1.7) Note if echelle voltage0 has changed. This should go into spreadsheet, too.
2.3) EXES has to be about 21 feet from SCHP.
2.extra) linearization, dark tests, and flat repeats are good "set and forget" type uses, particularly if we can make scripts.
8.4) the goal is to get slit 3 R>= 100,000 at C2H2. If we make that, don't worry about a better diagonal.
9+) the corner cube efficiency is worth checking. I'd suggest taking a low resolution gas cell H2O line at a known angle and checking to see what ADU count and line shape we get at same line in corner cube.
Thanks.
1.5) We use imaging mode normally, and we will do that first. But as an extra, I'm going to take a pupil image too, since we noticed the cross was visible in pupil mode in CD14. Not sure it will be useful, but its a fast extra.
2.3) Just meant as a starting point. And then we'll have to slowly back away, trying to keep the SCHP in the FOV.